Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 5 de 5
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Biomed Pharmacother ; 156: 113994, 2022 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-36411655

RESUMO

Organic Anion-Transporting Polypeptides (OATPs) are known to control the liver uptake of many drugs. Non-hepatic expression of OATPs has been reported although functional importance for whole-body pharmacokinetics (WBPK) remains unknown. Glyburide is a well described substrate of several hepatic and non-hepatic OATPs. Dynamic whole-body positron emission tomography (DWB-PET) with [11C]glyburide was performed in humans for determination of the importance of OATPs for liver uptake and WBPK. Seven healthy male subjects (24.7 ± 3.2 years) underwent [11C]glyburide PET scan with concomitant blood sampling. All subjects underwent baseline [11C]glyburide PET scan. Five subjects underwent a subsequent [11C]glyburide PET scan after infusion of the potent OATP inhibitor rifampicin (9 mg/kg i.v.). The transfer constant (kuptake) of [11C]glyburide from blood to the liver was estimated using the integration plot method. The tissue exposure of [11C]glyburide was described by the area under the time-activity curve (AUC) and corresponding tissue/blood ratio (AUCR). [11C]glyburide was barely metabolized in both the baseline and rifampicin conditions. Parent (unmetabolized) [11C]glyburide accounted for > 90 % of the plasma radioactivity. Excellent correlation was found between radioactive counting in arterial blood samples and in the image-derived input function, in both the baseline and rifampicin conditions (R2 = 97.9 %, p < 0.01). [11C]glyburide predominantly accumulated in the liver. Rifampicin decreased liver kuptake by 77.3 ± 7.3 %, which increased exposure in blood, kidneys, spleen, myocardium and brain (p < 0.05). No significant change in AUCR was observed except in the liver (p < 0.01). [11C]glyburide benefits from metabolic stability and high sensitivity to OATP inhibition which enables quantitative determination of OATP function. DWB-PET suggests negligible role for non-hepatic OATPs in controlling the tissue distribution of [11C]glyburide.


Assuntos
Glibureto , Transportadores de Ânions Orgânicos , Humanos , Masculino , Rifampina/farmacologia , Fígado/diagnóstico por imagem , Proteínas de Membrana Transportadoras , Tomografia por Emissão de Pósitrons , Peptídeos , Ânions
2.
Phys Med Biol ; 67(9)2022 04 27.
Artigo em Inglês | MEDLINE | ID: mdl-35395657

RESUMO

Objective.In clinical positron emission tomography (PET) imaging, quantification of radiotracer uptake in tumours is often performed using semi-quantitative measurements such as the standardised uptake value (SUV). For small objects, the accuracy of SUV estimates is limited by the noise properties of PET images and the partial volume effect. There is need for methods that provide more accurate and reproducible quantification of radiotracer uptake.Approach.In this work, we present a deep learning approach with the aim of improving quantification of lung tumour radiotracer uptake and tumour shape definition. A set of simulated tumours, assigned with 'ground truth' radiotracer distributions, are used to generate realistic PET raw data which are then reconstructed into PET images. In this work, the ground truth images are generated by placing simulated tumours characterised by different sizes and activity distributions in the left lung of an anthropomorphic phantom. These images are then used as input to an analytical simulator to simulate realistic raw PET data. The PET images reconstructed from the simulated raw data and the corresponding ground truth images are used to train a 3D convolutional neural network.Results.When tested on an unseen set of reconstructed PET phantom images, the network yields improved estimates of the corresponding ground truth. The same network is then applied to reconstructed PET data generated with different point spread functions. Overall the network is able to recover better defined tumour shapes and improved estimates of tumour maximum and median activities.Significance.Our results suggest that the proposed approach, trained on data simulated with one scanner geometry, has the potential to restore PET data acquired with different scanners.


Assuntos
Aprendizado Profundo , Neoplasias Pulmonares , Humanos , Processamento de Imagem Assistida por Computador/métodos , Neoplasias Pulmonares/diagnóstico por imagem , Imagens de Fantasmas , Tomografia por Emissão de Pósitrons
3.
Phys Med Biol ; 66(18)2021 09 16.
Artigo em Inglês | MEDLINE | ID: mdl-34433155

RESUMO

Dynamic whole body (DWB) PET acquisition protocols enable the use of whole body parametric imaging for clinical applications. In FDG imaging, accurate parametric images of PatlakKican be complementary to regular standardised uptake value images and improve on current applications or enable new ones. In this study we consider DWB protocols implemented on clinical scanners with a limited axial field of view with the use of multiple whole body sweeps. These protocols result in temporal gaps in the dynamic data which produce noisier and potentially more biased parametric images, compared to single bed (SB) dynamic protocols. Dynamic reconstruction using the Patlak model has been previously proposed to overcome these limits and shown improved DWB parametric images ofKi. In this work, we propose and make use of a spectral analysis based model for dynamic reconstruction and parametric imaging of PatlakKi. Both dynamic reconstruction methods were evaluated for DWB FDG protocols and compared against 3D reconstruction based parametric imaging from SB dynamic protocols. This work was conducted on simulated data and results were tested against real FDG dynamic data. We showed that dynamic reconstruction can achieve levels of parametric image noise and bias comparable to 3D reconstruction in SB dynamic studies, with the spectral model offering additional flexibility and further reduction of image noise. Comparisons were also made between step and shoot and continuous bed motion (CBM) protocols, which showed that CBM can achieve lower parametric image noise due to reduced acquisition temporal gaps. Finally, our results showed that dynamic reconstruction improved VOI parametric mean estimates but did not result to fully converged values before resulting in undesirable levels of noise. Additional regularisation methods need to be considered for DWB protocols to ensure both accurate quantification and acceptable noise levels for clinical applications.


Assuntos
Processamento de Imagem Assistida por Computador , Tomografia por Emissão de Pósitrons , Imagens de Fantasmas , Imagem Corporal Total
4.
J Nucl Med ; 62(6): 871-879, 2021 06 01.
Artigo em Inglês | MEDLINE | ID: mdl-33246982

RESUMO

This work set out to develop a motion-correction approach aided by conditional generative adversarial network (cGAN) methodology that allows reliable, data-driven determination of involuntary subject motion during dynamic 18F-FDG brain studies. Methods: Ten healthy volunteers (5 men/5 women; mean age ± SD, 27 ± 7 y; weight, 70 ± 10 kg) underwent a test-retest 18F-FDG PET/MRI examination of the brain (n = 20). The imaging protocol consisted of a 60-min PET list-mode acquisition contemporaneously acquired with MRI, including MR navigators and a 3-dimensional time-of-flight MR angiography sequence. Arterial blood samples were collected as a reference standard representing the arterial input function (AIF). Training of the cGAN was performed using 70% of the total datasets (n = 16, randomly chosen), which was corrected for motion using MR navigators. The resulting cGAN mappings (between individual frames and the reference frame [55-60 min after injection]) were then applied to the test dataset (remaining 30%, n = 6), producing artificially generated low-noise images from early high-noise PET frames. These low-noise images were then coregistered to the reference frame, yielding 3-dimensional motion vectors. Performance of cGAN-aided motion correction was assessed by comparing the image-derived input function (IDIF) extracted from a cGAN-aided motion-corrected dynamic sequence with the AIF based on the areas under the curves (AUCs). Moreover, clinical relevance was assessed through direct comparison of the average cerebral metabolic rates of glucose (CMRGlc) values in gray matter calculated using the AIF and the IDIF. Results: The absolute percentage difference between AUCs derived using the motion-corrected IDIF and the AIF was (1.2% + 0.9%). The gray matter CMRGlc values determined using these 2 input functions differed by less than 5% (2.4% + 1.7%). Conclusion: A fully automated data-driven motion-compensation approach was established and tested for 18F-FDG PET brain imaging. cGAN-aided motion correction enables the translation of noninvasive clinical absolute quantification from PET/MR to PET/CT by allowing the accurate determination of motion vectors from the PET data itself.


Assuntos
Encéfalo/diagnóstico por imagem , Fluordesoxiglucose F18 , Processamento de Imagem Assistida por Computador/métodos , Movimento , Redes Neurais de Computação , Tomografia por Emissão de Pósitrons , Humanos , Imageamento por Ressonância Magnética
5.
EJNMMI Phys ; 4(1): 31, 2017 Dec 11.
Artigo em Inglês | MEDLINE | ID: mdl-29230607

RESUMO

BACKGROUND: PET-MRI is under investigation as a new strategy for quantitative myocardial perfusion imaging. Consideration is required as to the maximum scanner count rate in order to limit dead-time losses resulting from administered activity in the scanner field of view during the first pass of the radiotracer. RESULTS: We performed a decaying-source experiment to investigate the high count-rate performance of a PET-MR system (Siemens mMR) over the expected range of activities during a clinical study. We also performed imaging of a cardiac perfusion phantom, which provides an experimental simulation of clinical transit of a simultaneous radiotracer (phantom injected activities range 252 to 997 MBq) and gadolinium-based contrast agent (GBCA). Time-activity and time-intensity curves of the aorta and myocardium compartments from PET and MR images were determined, and quantification of perfusion was then performed using a standard cardiac kinetic model. The decaying-source experiment showed a maximum noise equivalent count rate (NECRmax) of 286 kcps at a singles rate of 47.1 Mcps. NECR was maintained within 5% (NECR95%) of the NECRmax with a singles rate of 34.1 Mcps, corresponding to 310 MBq in the phantom. Count-rate performance was degraded above the singles rate of 64.9 Mcps due to the number of detection events impacting the quantitative accuracy of reconstructed images. A 10% bias in image activity concentration was observed between singles rates of 78.2 and 82.9 Mcps. Perfusion phantom experiments showed that image-based activity concentration and quantified values of perfusion were affected by count losses when the total singles rate was greater than 64.9 Mcps. This occurred during the peak arterial input function (AIF) phase of imaging for injected activities to the phantom of 600 MBq and greater. CONCLUSIONS: Care should be taken to avoid high count-rate losses in simultaneous PET-MRI studies. Based on our results in phantoms, bias in reconstructed images should be avoided by adhering to a singles rate lower than 64.9 Mcps on the mMR system. Quantification of perfusion values using singles rates higher than 64.9 Mcps on this system may be compromised and should be avoided.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...